Sterilization equipment for clinics that reprocess on site
Steam Sterilizer in a clinic reprocessing room

Novo Pro 45L Plus Steam Sterilizer

The Novo Pro 45L Plus is a Class B steam sterilizer built for high-volume reprocessing in clinics and surgical centers that handle large instrument loads.

Overview

What The Novo Pro 45L Plus Is For

The Novo Pro 45L Plus is a Class B steam sterilizer built for high-volume reprocessing in clinics and surgical centers that handle large instrument loads. A 45 L chamber accommodates up to five trays or three trays rotated 90 degrees, supporting multiple wrapped and hollow instrument sets in a single cycle. It runs on single-phase 220-240 V and occupies 685 x 438 mm of bench space, fitting into compact reprocessing rooms. Every cycle is monitored and recorded for compliance, and the intelligent system reduces cycle time through preheating and automatic phase management.

100+Countries with users
8Patented technologies
1800Days of aging test
24Months warranty
Key Benefits

Why Clinics Choose The Novo Pro 45L Plus

45 L chamber supports large instrument loads, reducing the number of cycles needed per session

Class B fractionated vacuum cycles process wrapped, hollow, and porous loads, not just unwrapped solids

Preheating with standby maintains chamber temperature between cycles, reducing start-up time

Intelligent monitoring and diagnosis identifies failed components and displays fault messages on screen

Every cycle is traceable via SD card or thermal printer, supporting compliance audits

Large wind cooling window with 73 aluminium layers and 18 pipes cools steam to water in seconds

Application Scenarios

  • Sterilizing multiple wrapped surgical kits and instrument sets in a single cycle for a busy dental or veterinary surgical center
  • Processing hollow instruments and porous items such as cotton packs in a single load with full vacuum penetration
  • Running high-volume end-of-day sterilization for a clinic with multiple chairs and rotating staff
  • Maintaining compliance records for inspections using automated cycle data stored on SD card or printed
  • Using the preheating system to reduce cycle time during peak hours when the schedule is full

Who It Is For

  • Dental clinics and stomatology departments with multiple chairs and high instrument turnover
  • Surgical centers and outpatient departments that sterilize large instrument sets in-house
  • Veterinary practices handling surgical packs and multiple hollow instruments
  • Aesthetic and cosmetic clinics that process reusable cannulas and instrument sets in bulk
  • Dental equipment distributors seeking a Class B model with full traceability and compliance features
Problems It Solves

What Goes Wrong Without It

  • Small sterilizers cannot process large wrapped or hollow loads, forcing multiple cycles and delays
  • Manual cycle tracking leads to lost records and audit failures during inspections
  • Cold starts extend cycle time at the beginning of the day, slowing down patient turnover
  • Inadequate cooling delays the next cycle and increases downtime between loads
  • Lack of diagnostic feedback makes troubleshooting difficult and extends repair time
Joident Novo Pro 45L Plus steam sterilizer, view 6
Technical

Technical Features And Specifications

  • Class B fractionated vacuum cycles conforming to EN 13060 and IEC 61010-2-040
  • 11 pre-set programs and tests with fully automatic phase management via microprocessor
  • Preheating system with standby to maintain minimum chamber temperature between cycles
  • Wind cooling super window with 260 x 142 mm cooling area, 73 aluminium radiating layers, and 18 radiating pipes
  • Intelligent monitoring and diagnosis system that identifies failure sections and displays fault messages
  • SD card or thermal printer (built-in or external) for cycle traceability and compliance records
  • Double safety door locking combining mechanical lock, sensor switch, and pressure lock
  • Manual or optional automatic water filling with level sensor to stop filling when tank is full
  • Built-in water quality detection as an optional feature to monitor feed water conductivity
  • Steel wire carrier and aluminium trays for efficient load organization and stability
Novo Pro 45L Plus specifications
Cycle TypeClass B, fractionated vacuum
Chamber Volume45 L
Usable Chamber Volume45 L
Electrical Supply220-240 Vac, 50/60 Hz, single phase
Nominal Voltage220 V, 50 Hz
Max Current10 A
Max Working Pressure2.5 bar (250 kPa)
Min Working Pressure-0.9 bar (-90 kPa)
Min Atmospheric Pressure0.7 bar
Max Temperaturebelow 145 C
Sound Power Levelbelow 80 dB
Water Qualityconductivity below 15 microsiemens per cm
Clean Water Tank3.5 L
Waste Water Tank1.7 L
Tray Capacityup to 5 trays, or 3 trays rotated 90 degrees
Bacteriological Filternot less than 99.5 percent for particles above 0.3 micrometres
Max Altitude2000 m above sea level
ConformityEN 13060, IEC 61010-1, IEC 61010-2-040, EN 61326, 93/42/EEC

Materials

  • Stainless steel sterilization chamber
  • Steel wire tray carrier
  • Aluminium instrument trays
  • Aluminium finned heat exchanger for the cooling window

How It Is Used In The Workflow

  • Load preparation - instruments are cleaned in an ultrasonic cleaner and dried before wrapping
  • Wrapping - pouches or reels are sealed on a thermal sealer with the cycle date visible
  • Water supply - the clean water tank is filled with distilled water below 15 microsiemens per cm
  • Cycle selection - the operator selects one of the 11 pre-set programs on the touch interface
  • Fractionated vacuum - repeated vacuum and steam pulses remove air from wrapped and hollow loads
  • Sterilization and drying - the cycle holds temperature, then dries, with extra drying available
  • Record - the cycle is written to SD card or printed via built-in or external printer

Quality Control

  • Vacuum and pressure limits are monitored on every cycle and the cycle aborts if a limit is missed
  • Programs and tests on board include the routine tests a practice runs before the first load of the day
  • The intelligent diagnosis system identifies the failing section rather than showing a generic error
  • Each unit is aged tested before shipment, with an aging test programme measured in days not hours
  • Cycle data is retained on removable media so a failed cycle can be traced after the fact

Certifications And Conformity

  • EN 13060 small steam sterilizer
  • IEC 61010-1
  • IEC 61010-2-040
Options

Configuration And Customization

Automatic water filling system in place of manual filling

Built-in water quality detection

Built-in or external thermal printer for cycle records

Instruments management assistant trays for mirrors, probes and similar small instruments

Multi language operating system set to the destination market

OEM and ODM builds for distributors, discussed case by case

Comparison

How The Novo Pro 45L Plus Compares

Compared WithWhat Changes
Class N sterilizerA Class N unit only handles unwrapped solid instruments; the Novo Pro 45L Plus also processes wrapped and hollow loads, at a higher purchase price
Novo Pro 23LThe 23L chamber suits a single to two chair practice and uses less water per cycle; the 45L takes more per load when the schedule is full
Novo Pro 45LThe 45L Plus includes an enhanced diagnostic system, preheating with standby, and a larger cooling window; the base 45L lacks these features
Central sterile servicesSending instruments out removes the equipment cost but adds a transport cycle and a second set of instruments
Use Guide

Running The Novo Pro 45L Plus Day To Day

  • Fill the clean water tank with distilled water below 15 microsiemens per cm and check the level indicator
  • Run the daily test programme before the first patient load
  • Load trays so pouches lie flat with the paper side up and do not touch the chamber wall
  • Select the program that matches the load: wrapped, hollow, porous, or mixed
  • Use extra drying for dense instrument assemblies or anything that has come out damp before
  • Empty the waste water tank when prompted and never return waste water to the clean tank
  • Keep the SD card in place so every cycle is recorded, and archive the file periodically
Joident Novo Pro 45L Plus steam sterilizer, view 7
FAQ

Novo Pro 45L Plus Questions

What is the difference between Class B and Class N sterilization
Class B uses fractionated vacuum to pull air out of the chamber before steam enters, so steam reaches wrapped instruments and the inside of hollow handpieces. Class N handles unwrapped solid instruments only. The Novo Pro 45L Plus runs Class B cycles conforming to EN 13060.
What water does the sterilizer need
Feed water must have conductivity below 15 microsiemens per centimetre. Distilled water from a Distilo water distiller meets this. Tap water scales the boiler and shortens the life of the machine.
How are cycle records kept for an inspection
Every cycle is written to SD card or printed via built-in or external thermal printer, so a paper or digital record can be attached to the day sheet.
What power supply does it need
220-240 Vac, 50/60 Hz single phase, nominally 220 V at 50 Hz, drawing up to 10 A. Plan a dedicated circuit rather than sharing a socket with other equipment.
Can it be used at altitude
The unit is rated to 2000 m above sea level. Above that height it must not be used, because the pressure and temperature relationship the cycle depends on no longer holds.
Do you supply OEM or ODM versions for distributors
Yes. Joident supplies distributors and OEM or ODM partners, including a multi language operating system set for the destination market. Send your requirement on WhatsApp or by email and we will confirm what is feasible.

Related Products

Joident Novo Pro 23L
Steam Sterilizer

Novo Pro 23L

The Novo Pro 23L is a Class B steam sterilizer designed for small to mid-size reprocessing rooms where space i.

Ask About The Novo Pro 45L Plus

Send the model name and your country and we will come back with a quote, lead time and the configuration options for the Novo Pro 45L Plus.