Sterilization equipment for clinics that reprocess on site
Joident Home

Specifying A Sterilization Room For A Dental Practice

The reprocessing room is usually the smallest room in the practice and the one an inspector looks at first. This is how to size the equipment for the chairs you actually run, and lay it out so instruments only travel one way.

Stage By Stage

Decisions In The Order You Have To Make Them

Get the sequence right and the room fits. Get it wrong and you buy a chamber that is too small in year two.

Stage 1

Map The Flow

Problem
Instruments that cross back over their own path undo the whole process.

How Joident Solves It
Lay the bench out one way, dirty intake to sterile storage, and place the four machines in that order.

Meto 30
Stage 2

Size The Chamber

Problem
An undersized chamber creates a queue on exactly the afternoon you cannot afford one.

How Joident Solves It
Size on the busiest session. Joident chambers run from 8 L to 45 L on the same Class B platform.

Distilo
Stage 3

Plan The Services

Problem
A sterilizer sharing a socket with the compressor trips the circuit.

How Joident Solves It
Novo Pro models run on single phase 200-240 Vac and a 23 L draws 2700 W with a 16 A fuse, so plan a dedicated circuit.

Seal
Stage 4

Set Up The Record

Problem
Handwritten logs are the first thing an inspection questions.

How Joident Solves It
Cycle data goes to SD card or USB automatically, with a built in or external printer where a paper copy is expected.

Novo Pro 45L Plus
Outcome

What A Well Specified Room Delivers

  • Instruments move one way, from dirty intake to sterile storage
  • The chamber matches the busiest session rather than the average day
  • Cycle records are produced as the machine runs
  • Staff training takes a shift, not a month
  • The room passes inspection on the evidence it already produces
Joident About

Equipment In This Solution

Feed Water Is Why Sterilizers Fail Early
Water Distiller

Distilo

The Distilo water distiller produces low-conductivity water for sterilization and instrument reprocessing in s.

Pouch Sealing And Cycle Traceability
Thermal Sealer

Seal

The Seal thermal sealer is designed for clinics that need reliable, consistent sealing of sterilization pouche.

Ultrasonic Cleaning Before Sterilization
Ultrasonic Cleaner

Meto 30

The Meto 30 is a 30 L medical ultrasonic cleaner designed for high-volume instrument cleaning in hospitals, su.

FAQ

Setting Up A Dental Sterilization Room Questions

How big a chamber does a four chair practice need
It depends on how many surgical packs you run rather than how many chairs you have. A four chair general practice usually settles on 23 L; a practice doing regular implant surgery moves up to 29 L or 45 L so a full pack fits in one cycle.
Should the ultrasonic cleaner be in the same room
Yes, at the dirty end. The point of a one way layout is that an instrument never crosses back over its own path, so cleaning sits at the intake side and sterile storage at the far end.
Do I need a printer on the autoclave
Not necessarily. Every Novo Pro writes cycle data to SD card or USB, which is usually enough. A built in or external printer is worth adding where local practice is to attach a paper record to the day sheet.
What power supply should I plan for
Novo Pro sterilizers run on single phase 200-240 Vac at 50/60 Hz and draw up to 2700 W, so plan a dedicated circuit with a 16 A protection rather than sharing a socket with the compressor.
Can you help with the layout before we build
Yes. Send the room dimensions and door positions and we will place the four footprints against them and flag anything that will not open properly.

Have Your Room Sized Before You Order

Send the room dimensions and your chair count on WhatsApp and we will come back with a layout and the models that fit it.