HMEF Or Bacterial And Viral Filter Which One Belongs In Your Circuit
Both filters protect a breathing circuit. Only one returns moisture. Choosing the wrong one either dries the airway or duplicates a humidifier you already have.
The Single Question That Decides It
Does the circuit already have active humidification? If it does, a bacterial and viral filter is what you need, because the HMEF would be duplicating a job already done. If it does not, the HMEF earns its place by returning moisture that would otherwise be lost, at 31 mg H2O per litre at a tidal volume of 500 ml on the adult type.
Dead Space Is The Paediatric Constraint
Dead space matters most on small tidal volumes. The adult HMEF carries 45 ml and covers 150 to 1500 ml tidal volumes. The paediatric type drops to 20 ml for 150 to 300 ml. The bacterial and viral filter sits at 32 ml across a 250 to 1500 ml range, which rules it out for the smallest patients.
Filtration Is Comparable Across Both
Both run BFE 99.9 to 99.999 percent and VFE 99 to 99.99 percent depending on grade, so filtration efficiency is rarely the deciding factor. Resistance differs slightly: the BV filter is under 1.2 cmH2O at 30 LPM and under 2.6 cmH2O at 60 LPM at the highest grade.
The Tracheostomy Case Is Different Again
A spontaneously breathing tracheostomy patient does not need a circuit filter, they need something light enough to hang off the tube. The tracheostomy filter weighs 7 g with 15 ml dead space and a 15F connector, with moisture loss of 9.0 mg H2O per litre at 500 ml tidal volume and a recommended tidal volume above 50 ml.
Do Not Combine An HMEF With An Active Humidifier
Putting a heat and moisture exchanger downstream of an active humidifier loads the filter element with condensate and drives resistance up. Pick one humidification strategy per circuit and specify the filter accordingly.
Products Discussed In This Guide

Heat Moisture Exchanger Filter
The Heat Moisture Exchanger Filter combines a breathing filter with moisture return, so inspired gas stays humidified…
View Product →
Disposable Bacterial and Viral Filter
The disposable bacterial and viral filter protects the breathing circuit and the patient from cross contamination,…
View Product →
Tracheostomy Filter
The tracheostomy filter is a heat and moisture exchanger made for tracheostomised patients breathing spontaneously…
View Product →
Spirometry Filter
The spirometry filter sits between the patient and a lung function machine or resuscitator to stop cross…
View Product →Questions On This Topic
Can I use an HMEF and an active humidifier together?
No. Use one humidification strategy per circuit. A heat and moisture exchanger downstream of an active humidifier loads the element with condensate.
Which filter suits a paediatric circuit?
The paediatric HMEF at 20 ml dead space, covering 150 to 300 ml tidal volumes.
What connections do these filters use?
22M/15F-22F/15M on the HMEF and 22M/15F-15M/22F on the bacterial and viral filter.