SRM-362 Standard Reference Meter(discontinued)
Function Features Specification Order Info Function SRM-362 can be used as standard meter to test electric meter and energy meter, and also can measure all kinds of AC electric quantity; Measure errors of single/three phase active and reactive energy meter,.
- Model
- SRM-362
- Product line
- Meter Test and Calibration
- Manufacturer
- Kongter Test & Measurement Co., Limited, Shenzhen, China
- Quality standard
- ISO 9001:2015
- Customisation
- Tailored configurations available on request
Send your application details straight to us on WhatsApp, WeChat or by email. We do not use enquiry forms.
Specification
Full Details for SRM-362
- SRM-362 can be used as standard meter to test electric meter and energy meter, and also can measure all kinds of AC electric quantity;
- Measure errors of single/three phase active and reactive energy meter, and standard deviation estimating value;
- Measure Voltage, Current, Active power, Reactive power, Apparent power, Phase angle, Power factor, frequency and so on;
- Measure distortion of Voltage and Current wave, 2nd~127th harmonics and inter-harmonics content, and disaplay harmonic bar graph and waveforms;
- Measure power stability;
- Real time display phasor diagram, easy to analyze wiring.
- 8 inch color touch screen, interface friendly, easy to operate
- Automatically switch to all Voltage and Current ranges, and no need to change wiring during measurement
- High measuring resolution, Voltage and Current resolution superior to 2×10-5 , phase resolution 0.001°, frequency resolution 0.01Hz
- Reactive power measurement includes 3P4W true reactive, 3P3W true reactive, 3 elements 90°artificial reactive, 2 elements 90°artificial reactive, 2 elements 60°artificial reactive etc. , which is suitable to test reactive power meter and reactive energy meter in various measuring principle
- Equipped with RS232, Ethernet, and WiFi interface,SRM-362 can either support stand-alone operation, PC control or handy wireless terminal control
- Automatically detect, diagnose and alarm for failure
- Remotely updating online, easily achieve software updating
- Output standard energy pulse, and the pulse constant can be set manually
- Support local calibration at users’ side
AC Voltage Measurement
Range: 75mV, 750mV, 7.5V, 30V, 100V, 300V, 660V automatically switch
Stability: ≤100ppm/year
Measurement accuracy: 0.05%RG
AC Current Measurement
Range: 500μA, 2mA, 10mA, 50mA, 200mA, 500mA, 2A, 5A, 10A, 30A, 120A automatically switch
Stability: ≤100ppm/year
Measurement accuracy: 0.05%RG
Power Measurement
Active/reactive power measurement accuracy: 0.05%RG
Phase angle measurement
Measuring range: 0°~360°
Resolution: 0.001°
Measurement accuracy: 0.05°
Frequency measurement
Measuring range: 40Hz ~70Hz
Resolution: 0.001Hz
Accuracy: 0.002 Hz
Harmonic Voltage
Harmonic order: 2nd ~127th
Harmonic amplitude: 0~100%
Harmonic accuracy: when Uh≥1%Un, 5% Uh; when Uh<1%Un, 0.05% Un.
Harmonic Current
Harmonic order: 2nd ~127th
Harmonic amplitude: 0~100%
Harmonic accuracy: when Uh≥3%Un, 5% Uh; when Uh<3%Un, 0.15% Un
Energy error measurement:
Active energy basic error limit:
±0.05%RD (Voltage 15V~660V, Current 0.05A~120A, PF≥0.5)
±0.1%RD (Voltage 15V~660V, Current 0.02A~0.05A, PF=1)
Reactive energy basic error limit:
±0.1%RD (Voltage 15V~660V, Current 0.05A~120A, PF≥0.5)
±0.2%RD (Voltage 15V~660V, Current 0.02A~0.05A, PF=1)
Energy pulse
Energy pulse output:
Pulse constant setting range: (1~599999999) r/kWh (r/kvarh)
Max. output frequency: 10KHZ
Energy pulse input:
Pulse constant setting range: (1~599999999) r/kWh (r/kvarh)
Calibration pulse setting range: 1~999999
Max. input frequency: 10KHZ
Other parameters:
Power supply: 90-265VAC/DC
Power frequency: 50Hz~60Hz
Power consumption: 100VA
Environment condition: 20℃~30℃,Humidity: RH≤85%
Storage environment: -20℃~50℃
Size: 400 (L)× 440(W)× 176mm(H)
Weight: 15kg
SRM-362 contains 2 models with different accuracy below:
SRM-362C, Accuracy Class: 0.05
SRM-362B, Accuracy Class: 0.1
Request A QuoteCustomers also viewed:
MTS-122 Meter Test SystemAccuracy class 0.1 & 0.05
MTS-100 Meter Test SystemMeter test system up to 100A
SPS-31 DC Standard SourceTest analog & digital meters
FAQ
Questions About This Model
Is this model certified
All Kongter products are strictly accredited by the international quality standard ISO 9001:2015.
Can this be customised for my application
Yes. Customised configurations are part of the standard offering. Tell us the system voltage, capacity or load you need to cover.
What is the lead time
Lead time depends on the configuration. Message us on WhatsApp at +86 135 1009 2292 or email [email protected] with your requirement and we will confirm it.
Do you provide training and support
Yes. Kongter's R&D team develops equipment aimed at field use, and product support is part of the offering. See the support pages for what is covered.
How do I get a quotation
Message us on WhatsApp or WeChat at +86 135 1009 2292, email [email protected], or call +86-755-2691 6832.
Related Models
More From Meter Test and Calibration








Request a Quote
Ask for a Price on SRM-362
Tell us the system you need to test and the quantity. We reply with a quotation and lead time.